Skip to product information
1 of 1

CRI

3 day rental of Colorimetry Research CR-100

3 day rental of Colorimetry Research CR-100

Regular price £250.00
Regular price Sale price £250.00
Sale Sold out

Schedule rental with uksupport@flandersscientific.com.

This listing is for a 3 day rental of the Colorimetry Research CR-100 colorimeter.

This rental is available for customers that own FSI monitors and will allow you to quickly perform AutoCal on DM and XMP series monitors.

The Colorimetry Research CR-100 is a high quality colorimeter designed specifically to work with virtually any display technology. The CR-100's machined aluminum and steel housing coupled with a completely solid state design ensure excellent durability while keeping the probe incredibly compact and lightweight (just 12 ounces / 340 grams). What sets the CR-100 apart from other colorimeters are its high quality tristimulus filters. Every CR-100 goes through a strict quality control process including measurement of every tristimulus filter with a double monochromator to ensure the following performance benchmarks are attained:

Guaranteed Typical
Y (Photopic) ≤ 2.0% 1.4 - 1.8%
Xr (Red) ≤ 3.0% 2.3 - 2.7%
Xb (Red under Blue) ≤ 1.2% 0.8 - 1.0%
Z (Blue) ≤ 3.5% 2.2 - 2.7%


The CR-100 can be used with the CRI Application for manual measurement and logging. It is also compatible with FSI's AutoCal, LightIllusion's LightSpace CMS*, LightIllusion's ColourSpace*, and Portrait Display's CalMan allowing for full 3D LUT based calibration of Flanders Scientific monitors.

The CR-100 can be used to measure a wide variety of display technologies and utilizes the Four-Color Matrix Method to calibrate the probe for use with specific display types/spectral distributions. The Four-Color Matrix Method, developed by the National Institute of Standards and Technology (NIST), improves the accuracy of tristimulus colorimeters for measurements of color displays and has been verified to be appropriate for cathode rays tubes (CRTs), liquid crystal displays (LCDs), and organic light emitting diodes (OLED) displays. When ordering the CR-100 (without a spectroradiometer) from Flanders Scientific the probe will be equipped with custom matrix matched selections for all current Flanders Scientific monitors in addition to any matrices provided by CRI.

*LightSpace CMS is sold separately.
*CalMAN is sold separately.
*ColourSpace is sold separately.

Wavelength Range
380 - 780 nm

Luminance Range
0.0002 fL to 1500 fL
 
0.0002 fL @ Signal to Noise Ratio > 10, 20 seconds exposure
 
0.001 fL @ Signal to Noise Ratio > 10, 0.5 second exposure
 
0.001 fL @ Signal to Noise Ratio > 100, 20 seconds exposure
 
1500 fL @ Signal to Noise Ratio > 8000, 1 second exposure

Luminance Accuracy
± 2 % @ 0.1 fL, 0.4 second exposure

Luminance Repeatability
0.2 % @ 0.1 fL, 0.4 second exposure
 
1.5 % @ 0.01 fL, 0.5 second exposure
 
1.5 % @ 0.001 fL, 20 seconds exposure

Chromaticity Accuracy
± 0.0015 x, y @ 0.1 fL, 0.4 second exposure
 
± 0.0015 x, y @ 0.02 fL, 20 seconds exposure

Chromaticity Repeatability
0.0005 x, y @ 0.1 fL, 0.4 second exposure
 
0.0005 x, y @ 0.02 fL, 20 seconds exposure

Polarization Error
≤ 0.1 %

Synchronization Frequencies
10 - 500 Hz (Custom Synchronization from 10 Hz to 10 KHz available)

Exposure Time Range
1 ms to 20 seconds

Power Requirements
5V, 120 mA (600 mW) via USB 2.0

Interface
USB 2.0, Ethernet

Weight
340 g
View full details